Lataa...

Single-chamber, rate-responsive pacemaker-mediated tachycardia

The case of a patient with a unipolar VVIR, who developed a sensor-driven increased pacing rate at rest following pulse generator replacement, is presented. The cause was pectoral muscle stimulation, which triggered a sensor-driven rate response in the supine position. The possible causes and manage...

Täydet tiedot

Tallennettuna:
Bibliografiset tiedot
Päätekijät: Bohm, Adam, Kiss, Robert, Dorian, Paul, Pinter, Arnold
Aineistotyyppi: Artigo
Kieli:Inglês
Julkaistu: Pulsus Group Inc 2010
Aiheet:
Linkit:https://ncbi.nlm.nih.gov/pmc/articles/PMC2989359/
https://ncbi.nlm.nih.gov/pubmed/21076726
Tagit: Lisää tagi
Ei tageja, Lisää ensimmäinen tagi!