Učitavanje...
Interruption in the supply of breathing gas during general anesthesia due to malposition of the vaporizer -A case report-
We report a case of interruption in the supply of breathing gas during general anesthesia caused by malposition of the Drager Vapor 2000® vaporizer, which was accidentally tilted and lifted off the Selectatec manifold of the anesthesia machine. Because the patient was an 1-month-old infant, we could...
Spremljeno u:
Glavni autori: | , |
---|---|
Format: | Artigo |
Jezik: | Inglês |
Izdano: |
The Korean Society of Anesthesiologists
2010
|
Teme: | |
Online pristup: | https://ncbi.nlm.nih.gov/pmc/articles/PMC2966709/ https://ncbi.nlm.nih.gov/pubmed/21057618 https://ncbi.nlm.nih.govhttp://dx.doi.org/10.4097/kjae.2010.59.4.270 |
Oznake: |