A fast cardiac gamma camera with dynamic SPECT capabilities: design, system validation and future potential

PURPOSE: The goal of this study is to present the Discovery NM 530c (DNM), a cardiac SPECT camera, interfacing multi-pinhole collimators with solid-state modules, aiming at slashing acquisition time without jeopardizing quality. DNM resembles PET since it enables 3-D SPECT without detector motion. W...

Celý popis

Uloženo v:
Podrobná bibliografie
Hlavní autoři: Bocher, Moshe, Blevis, Ira M., Tsukerman, Leonid, Shrem, Yigal, Kovalski, Gil, Volokh, Lana
Médium: Artigo
Jazyk:Inglês
Vydáno: Springer-Verlag 2010
Témata:
On-line přístup:https://ncbi.nlm.nih.gov/pmc/articles/PMC2933031/
https://ncbi.nlm.nih.gov/pubmed/20585775
https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1007/s00259-010-1488-z