NEMA NU 2-2007 performance measurements of the Siemens Inveon(™) preclinical small animal PET system
National Electrical Manufacturers Association (NEMA) NU 2-2007 performance measurements were conducted on the Inveon(™) preclinical small animal PET system developed by Siemens Medical Solutions. The scanner uses 1.51 × 1.51 × 10 mm LSO crystals grouped in 20 × 20 blocks; a tapered light guide coupl...