A Filtered Backprojection Algorithm for Triple-Source Helical Cone-Beam CT
Multisource cone-beam computed tomography (CT) is an attractive approach of choice for superior temporal resolution, which is critically important for cardiac imaging and contrast enhanced studies. In this paper, we present a filtered-backprojection (FBP) algorithm for triple-source helical cone-bea...