Mechanical Properties, Oxidation, and Clinical Performance of Retrieved Highly Crosslinked Crossfire Liners after Intermediate-Term Implantation
Sixty Crossfire (Stryker Orthopaedics, Mahwah, NJ) liners were consecutively revised after an average of 2.9 years (range: 0.01 – 8.0 years) for reasons unrelated to wear or mechanical performance of the polyethylene. Femoral head penetration was measured directly from 42 retrievals implanted for ov...