Monte Carlo simulation of a multi-leaf collimator design for telecobalt machine using BEAMnrc code

This investigation aims to design a practical multi-leaf collimator (MLC) system for the cobalt teletherapy machine and check its radiation properties using the Monte Carlo (MC) method. The cobalt machine was modeled using the BEAMnrc Omega-Beam MC system, which could be freely downloaded from the w...