A rare case of suture material obstructing the closure mechanism of a prosthetic aortic valve: a case report

Prosthetic aortic valve dysfunction presenting as aortic regurgitation is a complication of mechanical valve replacement. We describe a case of late valve dysfunction caused by an annular suture of excessive length obstructing the closure mechanism of a bileaflet prosthetic valve. We present this ra...