Accuracy of the SEVEN® Continuous Glucose Monitoring System: Comparison with Frequently Sampled Venous Glucose Measurements
BACKGROUND: The purpose of this study was to compare the accuracy of measurements obtained from the DexCom™ SEVEN® system with Yellow Springs Instrument (YSI) laboratory measurements of venous blood glucose. METHODS: Seventy-two subjects with insulin-requiring diabetes, aged 18–71, were enrolled in...