Sequential activation of multiple grounding pads reduces skin heating during radiofrequency tumor ablation
PURPOSE: Radiofrequency (RF) tumor ablation has become an accepted treatment modality for tumors not amenable to surgery. Skin burns due to ground pad heating may become a limiting factor for further increase in ablation zone dimensions and generator power. We investigated a method were groups of gr...