Inappropriate Shock Delivered By Implantable Cardioverter Defibrillator - Cardiac Resynchronization Therapy (ICD-CRT) Due To Myopotential Oversensing

The clinical efficacy of ICD-CRT therapy depends on accurate sensing of intracardiac signals and sensing algorithms. We report the occurrence of sensing abnormality in a patient with ICD-CRT. In this patient, oversensing of myopotentials during strenuous muscular activity resulted in an inappropriat...