A Prototype RF Dosimeter for Independent Measurement of the Average Specific Absorption Rate (SAR) During MRI

PURPOSE: To develop a scanner-independent dosimeter for measuring the average radio frequency (RF) power deposition and specific absorption rates (SAR) for human MRI exposure. MATERIALS AND METHODS: A prototype dosimeter has a transducer with orthogonal conducting loops surrounding a small signal-ge...