Measured count-rate performance of the Discovery STE PET/CT scanner in 2D, 3D and partial collimation acquisition modes

We measured count rates and scatter fraction on the Discovery STE PET/CT scanner in conventional 2D and 3D acquisition modes, and in a partial collimation mode between 2D and 3D. As part of the evaluation of using partial collimation, we estimated global count rates using a scanner model that combin...

Полное описание

Сохранить в:
Библиографические подробности
Главные авторы: MacDonald, L R, Schmitz, R E, Alessio, A M, Wollenweber, S D, Stearns, C W, Ganin, A, Harrison, R L, Lewellen, T K, Kinahan, P E
Формат: Artigo
Язык:Inglês
Опубликовано: 2008
Предметы:
Online-ссылка:https://ncbi.nlm.nih.gov/pmc/articles/PMC2585489/
https://ncbi.nlm.nih.gov/pubmed/18574308
https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1088/0031-9155/53/14/002