Validation of an ion selective electrode system for the analysis of serum fluoride ion

A high impedance unit was developed for use with a fluoride/pH electrode system for the measurement of serum fluoride. The linearity, accuracy, precision and detection limit of the system is reported. At a pH of 1.55, the system was linear over a range of serum fluoride concentrations up to 100 μmol...