The use of spore strips for monitoring the sterilization of bottled fluids.

A bacterial spore test has been developed which enables the efficacy of the sterilizing cycle recommended by the British Pharmaceutical Codex (1973) for bottled fluids to be accurately monitored. During a 14-month period this test detected faults in 3.3% of the sterilizing cycles, representing five...

Celý popis

Uloženo v:
Podrobná bibliografie
Hlavní autoři: Selkon, J. B., Sisson, P. R., Ingham, H. R.
Médium: Artigo
Jazyk:Inglês
Vydáno: Cambridge University Press 1979
Témata:
On-line přístup:https://ncbi.nlm.nih.gov/pmc/articles/PMC2130100/
https://ncbi.nlm.nih.gov/pubmed/458140