Chargement en cours...
Variance in multiplex suspension array assays: carryover of microspheres between sample wells
BACKGROUND: This study was undertaken because of the accidental observation that a sample of 60+ beads was obtained by the instrument from a completely dry, unused well in a 96 well plate. Others have observed unexplained outliers in replicated wells. The problem was first observed on an older instr...