Performance characteristics of bioassay, radioenzymatic assay, homogeneous enzyme immunoassay, and high-performance liquid chromatographic determination of serum gentamicin.
We compared the accuracy, precision, and between-method error of the microbiological assay, the radioenzymatic assay, the homogeneous enzyme immunoassay, and the high-performance liquid chromatographic assay for the quantitation of gentamicin in serum. Precision and accuracy were evaluated by refere...