Transcutaneous estimation of arterial carbon dioxide in intensive care. Which electrode temperature?

Transcutaneous and arterial carbon dioxide were measured simultaneously in 57 children (age range 10 days to 14.3 years) undergoing intensive care. All were haemodynamically stable at the time of study. Mean calibration time with 5 and 10% carbon dioxide was 43 (range 38-58) minutes and mean arteria...