A novel approach to temporary stenting: degradable cardiovascular stents produced from corrodible metal—results 6-18 months after implantation into New Zealand white rabbits

OBJECTIVE—To determine whether corrodible materials may be safely used as biodegradable cardiovascular implants.
DESIGN—Corrodible iron stents (> 99.8% iron) were produced from pure iron and laser cut with a stent design similar to a commercially available permanent stent (PUVA-AS16). A total of...

ver descrição completa

Na minha lista:
Detalhes bibliográficos
Main Authors: Peuster, M, Wohlsein, P, Brugmann, M, Ehlerding, M, Seidler, K, Fink, C, Brauer, H, Fischer, A, Hausdorf, G
Formato: Artigo
Idioma:Inglês
Publicado em: 2001
Assuntos:
Acesso em linha:https://ncbi.nlm.nih.gov/pmc/articles/PMC1729971/
https://ncbi.nlm.nih.gov/pubmed/11602554