A carregar...
Aortic obstruction caused by device occlusion of patent arterial duct
A 2 year old girl is reported in whom deployment of the Amplatzer ductal occluder caused significant aortic obstruction, requiring surgical removal of the device. This case emphasises the need for careful echocardiographic and angiographic assessment of the position of the Amplatzer ductal occluder...
Na minha lista:
| Main Authors: | , |
|---|---|
| Formato: | Artigo |
| Idioma: | Inglês |
| Publicado em: |
1999
|
| Assuntos: | |
| Acesso em linha: | https://ncbi.nlm.nih.gov/pmc/articles/PMC1729088/ https://ncbi.nlm.nih.gov/pubmed/10377324 |
| Tags: |
Adicionar Tag
Sem tags, seja o primeiro a adicionar uma tag!
|