Transthoracic ventricular defibrillation in adults.
A prospective study of the energy required for transthoracic ventricular defibrillation in adults showed that in 42 (81%) out of 52 episodes of ventricular fibrillation shocks of 100 watt-seconds (Ws) of stored energy were successful. Out of 233 episodes, 222 (95%) were converted by 200 W s shocks....