Voltage clamp limitations of dual whole-cell gap junction current and voltage recordings. I. Conductance measurements.
Previous correction methods for series access resistance errors in the dual whole-cell configuration did not take into account the effect of nonzero resting potentials (E(rest)) and junctional reversal potentials (E(rev)). Dual whole-cell currents were modeled according to resistor-circuit analysis...