Surface Characterization of Retrieved Metal-on-Metal Total Hip Implants from Patients with Adverse Reaction to Metal Debris

The use of metal-on-metal (MoM) total hip implants has decreased recently due to reports of high failure rates and adverse local tissue reaction (ALTR). It has been hypothesized that wear metal debris released from CoCr bearing surfaces may provoke delayed hypersensitivity reactions. The goal of thi...

ver descrição completa

Na minha lista:
Detalhes bibliográficos
Main Authors: Maria Burbano, Robert Russell, Michael Huo, Robert Welch, Diana Roy, Danieli C. Rodrigues
Formato: Artigo
Idioma:Inglês
Publicado em: MDPI AG 2014-03-01
Colecção:Materials
Assuntos:
Acesso em linha:http://www.mdpi.com/1996-1944/7/3/1866