Voltammetric Determination of Tetracycline by Using Multi-Wall Carbon Nanotube – Ionic Liquid Film Coated Glassy Carbon Electrode
A voltammetric method was developed for the determination of tetracycline (TC) by using an ionic liquid (IL, 1–octyl–3–methylimidazolium–hexafluorophosphate) – multi-wall carbon nanotube (MWNT) film coated glassy carbon electrode (GCE). Experiment showed that both IL and MWNT could facilitate the TC...