Three corrections for overshoot effect improved the dose for step-and-shoot intensity-modulated radiation therapy.
We measured the overshoot effect in a linac and reduced it using block correction, reverse-sequence correction, and index correction. A StarTrack detector was used on a Varian iX. Five segments, 1 × 10 cm2 in area, were designed; the centers were at -4, -2, 0, 2, and 4 cm on the x axis for measuring...
Sparad:
| Huvudupphov: | , , , |
|---|---|
| Materialtyp: | Artigo |
| Språk: | Inglês |
| Utgiven: |
Public Library of Science (PLoS)
2021-01-01
|
| Serie: | PLoS ONE |
| Länkar: | https://doi.org/10.1371/journal.pone.0250243 |
| Taggar: |
Inga taggar, Lägg till första taggen!
|