Cole type Pitot tube, discharge factor survey and calibration
This article proposes that the Pitot Cole tube used for measuring water flow in the distribution does not require calibration, as it is an intrinsically self-calibrated equipment. Its function is to transform flow speed (kinetic energy) into manometric pressure, where the measurement is made by mano...
Sábháilte in:
| Príomhchruthaitheoirí: | , , |
|---|---|
| Formáid: | Artigo |
| Teanga: | Inglês |
| Foilsithe / Cruthaithe: |
Elsevier
2024-06-01
|
| Sraith: | Measurement: Sensors |
| Ábhair: | |
| Rochtain ar líne: | http://www.sciencedirect.com/science/article/pii/S2665917424001284 |
| Clibeanna: |