Analytical Evaluation of Point-of-Care Finecare™ Procalcitonin Rapid Quantitative Test in Sepsis Population as Compared with Elecsys<sup>®</sup> BRAHMS Procalcitonin Immunoassay

The study compared two plasma procalcitonin (PCT) assays, the point of care (POC) Finecare™ Procalcitonin Rapid Quantitative Test and the Elecsys<sup>®</sup> BRAHMS PCT immunoassay, in sepsis ICU patients. Forty-one plasma samples were analyzed, showing a strong correlation (r = 0.98) and no signifi...

ver descrição completa

Na minha lista:
Detalhes bibliográficos
Principais autores: Mohd Zulfakar Mazlan, Wan Norlina Wan Azman, Najib Majdi Yaacob, Tan Say Koon, Nurul Khaiza Yahya
Formato: Artigo
Idioma:Inglês
Publicado em: MDPI AG 2024-05-01
coleção:Diagnostics
Assuntos:
Acesso em linha:https://www.mdpi.com/2075-4418/14/11/1080