Delayed migration due to shortening of the lower part of AFX endograft's main body in angled fusiform abdominal aortic aneurysm
We encountered two fusiform abdominal aortic aneurysm cases with delayed AFX endograft (Endologix Inc) migration >4 years after placement. These cases showed shortening and slight angulation of the main body in the anteroposterior direction. We speculate that the potential mechanism relates to the A...