Non-contact tonometry in the postoperative eye.
The Keeler Pulsair non-contact tonometer was compared with the Goldmann applanation tonometer in a series of 48 eyes after operation. A correlation coefficient of 0.92 (p less than 0.001) was found between the two instruments, with the Pulsair having a statistically significant tendency to overread...