Lanean...
Non-contact tonometry in the postoperative eye.
The Keeler Pulsair non-contact tonometer was compared with the Goldmann applanation tonometer in a series of 48 eyes after operation. A correlation coefficient of 0.92 (p less than 0.001) was found between the two instruments, with the Pulsair having a statistically significant tendency to overread...
Gorde:
| Egile nagusia: | |
|---|---|
| Formatua: | Artigo |
| Hizkuntza: | Inglês |
| Argitaratua: |
1989
|
| Gaiak: | |
| Sarrera elektronikoa: | https://ncbi.nlm.nih.gov/pmc/articles/PMC1041707/ https://ncbi.nlm.nih.gov/pubmed/2713301 |
| Etiketak: |
Etiketa erantsi
Etiketarik gabe, Izan zaitez lehena erregistro honi etiketa jartzen!
|