Importance of exercise testing shortly after subcutaneous implantable cardioverter-defibrillator implantation in patients with Brugada syndrome – The first case of associated inappropriate shock in Japan
We report the case of a 51-year-old patient with Brugada syndrome (BrS) who experienced inappropriate shock due to T-wave oversensing (TWOS) during exercise when the optimal sensing vector was selected based on the automatic analysis by a subcutaneous implantable cardioverter-defibrillator (S-ICD)....